The SANPO DN20 Compact Electromagnetic Water Meter measures flow by electromagnetic induction, with the converter and display mounted directly on the sensor body. The DN20 configuration has a PTFE liner, a pressure rating of PN16, a maximum liquid temperature of 100°C, an RS485 signal and a flanged connection. It is a small size suited to DN20 pipelines that need electronic flow monitoring instead of a mechanical meter.
Introduction to the SANPO DN20 Compact Electromagnetic Water Meter
The SANPO DN20 meter is used on clean water lines carrying an electrically conductive liquid. There are no vanes or moving parts inside the measuring tube, so the liquid passes straight through the measuring section.
At DN20, the flow on the line is usually small and can vary a lot during the day. Before choosing a configuration, you need to know the normal, minimum and maximum flow of the line.
What is a compact (integral-display) design?
In the compact version, the signal converter and the display sit directly on top of the sensor body as one unit. The operator reads the value on site and does not need a cable between the sensor and a remote display. A remote version is better when the measuring point is hard to reach or the display must be placed elsewhere.
How does it differ from a mechanical "magnetic-drive" water meter?
A mechanical magnetic-drive meter has a rotating element driven by the water and a magnetic coupling to the register. An electromagnetic meter has no rotating part: it measures the voltage induced when a conductive liquid flows through a magnetic field. The two types differ completely in principle, installation and power requirements.
Technical specifications of the SANPO DN20 Compact Electromagnetic Water Meter
|
Specification |
Value |
|
Product |
SANPO DN20 Compact Electromagnetic Water Meter |
|
Brand |
SANPO |
|
Measuring principle |
Electromagnetic |
|
Display type |
Integral (compact) |
|
Size |
DN20 |
|
Medium |
Clean water, conductive liquid |
|
Liner |
PTFE |
|
Nominal pressure |
PN16 |
|
Maximum temperature |
100°C |
|
Communication signal |
RS485 |
|
Connection |
Flanged |
|
Verification / calibration |
Not included |
The flow range of the DN20 is not stated in the available documents, so no figure is given here. Minh Hoa Thanh confirms the flow range for the ordered configuration.
Structure of the SANPO DN20 Compact Electromagnetic Water Meter
Structure diagram of the SANPO compact electromagnetic flow meter
Sensor body and measuring tube
The measuring tube is the section the liquid flows through, inside the magnetic field. The bore is continuous and has no parts that obstruct the flow.
PTFE liner
PTFE is the electrically insulating liner on the inside of the measuring tube; it is not the material of the whole meter body. The liner keeps the electrode signal from leaking through the metal body. When installing, do not let tools or an off-centre gasket scratch the edge of the liner.
Magnetic coil and electrodes
The coil creates a magnetic field perpendicular to the flow. Two electrodes placed opposite each other pick up the induced voltage and send it to the converter.
Converter and integral display
The converter turns the electrode signal into instantaneous flow and total volume, shown on the display mounted on the meter body.
RS485 port and cable gland
The RS485 port sends data to a control cabinet or monitoring system. Cable glands must be tightened to keep the electrical housing sealed.
Two connection flanges
The two ends of the meter are flanges that bolt to the DN20 pipe flanges. The available documents do not give the flange standard, so check the outer diameter, bolt circle and number of bolts against the delivered unit before preparing the pipework.
Operating principle of the SANPO DN20 electromagnetic meter
Operating principle of the SANPO electromagnetic flow meter
The meter works according to Faraday's law of electromagnetic induction. When a conductive liquid flows through the magnetic field in the measuring tube, a voltage proportional to the flow velocity appears between the two electrodes. The converter multiplies the velocity by the tube cross-section to obtain the volumetric flow, then accumulates it into a total volume.
Because the measurement is electrical, the liquid must be conductive. An electromagnetic meter is not used for gas, steam, oil or demineralised water with very low conductivity.
Applications of the SANPO DN20 Compact Electromagnetic Water Meter
The DN20 size is common on branch lines and process lines with small flow, where electronic data is needed for remote monitoring.
Small process water lines
Measures clean water supplied to a machine, a mixing tank or a production step in a plant, to track consumption per shift.
Small filtration and water treatment systems
Monitors the flow of filtered water or make-up water in a treatment unit with DN20 piping.
Measuring points that connect to a monitoring system
Use the RS485 port to send data to a PLC, SCADA system or data logger instead of reading values by hand.
Conductive wastewater lines
Can be considered for wastewater lines with suitable conductivity. In that case, the liquid composition, solids and air bubbles at the measuring point must be clarified.
Installation guide for the SANPO DN20 electromagnetic meter
Install in the correct flow direction
Follow the arrow on the sensor body. If the installed direction differs from the actual flow, correct it through the device settings as described in the manual; do not swap cables or electrodes.
Keep the pipe full
Choose a position where the measuring tube is always full. Avoid the highest point of the line where air collects, and avoid a free discharge right after the meter.
Straight pipe before and after the meter
Provide straight pipe upstream and downstream according to the manufacturer's instructions for the exact model, away from pumps, throttling valves and elbows.
Mount the flanges and protect the liner
Align the two pipe flanges concentrically before placing the meter. Centre the gasket so it does not cover the measuring bore, and tighten the bolts in a cross pattern.
Grounding and wiring
Ground the meter according to its wiring diagram, especially when the pipe is made of a non-conductive material. Check that the supply voltage matches the ordered version before powering on.
Commissioning and checks
Let the liquid fill the pipe, check the flange joints for leaks, read the display and test RS485 communication with the receiving system. Record the initial reading at handover.
Contact Minh Hoa Thanh for the SANPO DN20 Compact Electromagnetic Water Meter
Minh Hoa Thanh supplies SANPO electromagnetic meters and KOMAX water meters, and helps you choose a configuration for the real pipeline. When you contact us, please prepare:
-
The type of liquid, and the temperature and pressure at the installation point.
-
The normal, minimum and maximum flow of the DN20 line.
-
The dimensions of the existing pipe flanges.
-
The data communication and power supply requirements at the site.
-
Any verification or calibration requirement for the project.
CÔNG TY TNHH TM DV XNK MINH HÒA THÀNH
Address: C62, KDC Thới An, Đường Lê Thị Riêng, Phường Thới An, TP.HCM
Hotline/Zalo: 0982.666.352
Phone: 0286.250.2299
Email: minhhoathanh@gmail.com
Frequently asked questions about the SANPO DN20 electromagnetic meter
What flow range can the SANPO DN20 electromagnetic meter measure?
The available documents do not state a flow range for the DN20. It is confirmed for the ordered configuration; please give the normal, minimum and maximum flow of your line so we can check it.
Is PTFE the material of the meter body?
No. PTFE is the electrically insulating liner inside the measuring tube. The body material and the electrode material should be checked against the documents of the delivered unit.
What communication signal does the meter have?
This configuration has RS485. The protocol, device address and baud rate must be agreed with the receiving system during commissioning.
Can an electromagnetic meter measure any liquid?
No. It only measures conductive liquids. Do not use it for gas, steam, oil or water with very low conductivity; the liquid temperature must not exceed 100°C.
What does PN16 mean?
PN16 is the nominal pressure rating of the meter body and flanges. Compare it with the actual pressure of the pipeline when designing the system.
Does the product include verification or calibration?
The product is supplied without verification or calibration. If the project requires it, please tell us before ordering.